Next: AAOmega Spectrograph Control
Up: The Control Task
Previous: CCD loader and CCD
  Contents
The ADC control window (shown in Figure
) contains a
display of the ADC prism angles and buttons to allow the ADC to be
nulled, stopped or set to track a specified telescope position. In
normal operation this window is not needed as the ADC can be
controlled from the telescope control window.
User interaction is typically limited to using the 5 buttons on the
bottom left. If the system has failed for some reason, these allow
the user to recover ADC operations by slewing the ADC to the desire
position. The options are:
- Stop the ADC -- This should be done at the end of the
night
- Set tracking -- To start the ADC tracking with the
telescope
- Null ADC -- Null both elements, remove the ADCs
corrective power
- Slew to Tel Pos -- Rapid slew the ADC to match
Telescope position
- Update Position -- never used
Figure:
The ADC Control Window
- This is obtained by clicking on the more button in the ADC
section of the main Control Task window
 |
Next: AAOmega Spectrograph Control
Up: The Control Task
Previous: CCD loader and CCD
  Contents
Rob Sharp
2006-10-30