next up previous contents
Next: Controlling the detector Up: Controlling the spectrograph Previous: Setting and Displaying

Pre--observing Checks

 


Focussing UCLES
The depth of focus for UCLES is substantial and it will be rare that focus changes are required, but it should nevertheless be checked.
Configure spectrograph to wavelength of interest.
CCD: use a narrow unbinned window ( e.g. TEK1K_SETUP or THOMSON_SETUP)
IPCS: use ND filters over the arc lamps ( i.e. LF1 8 and LF2 3) and gradually remove them until the maximum countrate is below 1 Hz. Because of the large IPCS window used with UCLES the countrate limitations are severe and it is advised that the countrate should not exceed 0.5 Hz in any of the measured lines.


Focussing UHRF
First focus on the red laser line at 6328.160 Å and inspect the instrumental profile. It should have a Gaussian FWHM of approximately 48 m, plus a low asymmetric tail to higher wavelengths. Tests with a green laser yielded the same value. Note that UHRF resolves the ThAr lines, so these cannot be used to assess the instrumental profile, but they are used to refocus UHRF for each wavelength setting, which is essential.
Configure spectrograph to wavelength of interest.
CCD: use a narrow binned window ( e.g. UHRF_X4)
IPCS: use ND filters over the arc lamps ( i.e. LF1 8 and LF2 3) and gradually remove them until the maximum countrate is below 1 Hz.

Note that the HP UP configuration produces brighter spectra than the HP DOWN configuration.

The geometric structure of the image slicer modulates the signal in the spatial direction. This is clearly visible in the UHRF laser, ThAr, and quartz lamp exposures (see Figs gif and gif). Observers may find it useful to take flat field exposures not only through the image slicer but also through the slit, using the maximum slit length ( SL MAX). These will provide more uniform illumination in the spatial direction which makes it easier to identify pixel--to--pixel variations. (Reminder: SLIT SLIT and SLIT SLICER swap between the image slicer and slit assembly.)

  
Figure: UHRF, 10 sec ThAr lamp @ 6328 Å, SLICER, UHRF_X4, XMEM orientation.

  
Figure: UHRF, 10 sec quartz lamp, SLICER, UHRF_X4, XMEM orientation.


Wavelength placement:
Even after doing a CONFIG, further adjustment of the spectrum positioning may be desired, especially with UHRF for which the spectrum coverage is so small, or for UCLES if partial orders are being recorded at one or both sides of the detector, in which case it might be desirable to shift the spectrum in the cross-dispersion direction to obtain a better coverage of one of these. With UHRF, observers may wish to force the VAX to use encoder units, rather than transformed units ( e.g. mm), by specifying RAW. To return to transformed units, specify REAL.

The most likely parameters to change are:

Once satisfactory grating and focus settings are obtained, they can be locked into the control files with


Other useful commands:



next up previous contents
Next: Controlling the detector Up: Controlling the spectrograph Previous: Setting and Displaying



Helen Davies
hdd@aaoepp.aao.gov.au