next up previous contents
Next: Online Data Analysis Up: Controlling the detector Previous: CCDs

IPCS

 

Check that desired IPCS pixel size has been set (typically 74 x 17 microns for UCLES). If not, adjust x-gain and y-gain potentiometers on IPCS rack in coudé anteroom; read graphs to get voltages which correspond to desired pixel sizes.

Check that IPCS rack is switched to control room.

Set xlines and ylines switches (in control room) to give required number of pixels. The IPCS is normally used with a hardware format of 2048 256, a software window of 2040 240 and starting values of 14 2.

Image display orientations

      UCLES CCD display on XMEM      UCLES CCD display in FIGARO
     ___________________________    _____________________________
     | red                :   o|    | blue                 :   o|
     |  :                 :   v|    |  :                   :   v|
     |  :                 :   e|    |  :                   :   e|
     |  :                 :   r|    |  :                   :   r|
     |  :  red .... blue  :   s|    |  :  red ...... blue  :   s|
     |  :                 :   c|    |  :                   :   c|
     |  :                 :   a|    |  :                   :   a|
     |  :                 :   n|    |  :                   :   n|
     | blue               :   .|    | red                  :   .|
     ---------------------------    -----------------------------


     UCLES IPCS display on 1311A     UCLES IPCS display in FIGARO
    _____________________________    _____________________________
    | blue                      |    | blue------------------red |
    |  :                        |    |           red             |
    |  :                        |    |            .              |
    |  :                        |    |            .              |
    |  :  blue . . . . . . red  |    |            .              |
    |  :                        |    |            .              |
    |  :                        |    |            .              |
    |  :                        |    |            .              |
    | red                       |    |           blue            |
    -----------------------------    -----------------------------


 UHRF CCD display on XMEM      
     _____________     
     | blue   : o|     Increasing UG moves spectrum ---->
     |  :     : v|     Tek width in UG  ~= 1200 at R = 0.3M, 600 at R = 0.6M,
     |  :     : e|                          400 at R = 1.0M
     |  :     : r|
     |  :     : s|     Increasing UT moves spectrum down
     |  :     : c|     Tek height in UT ~= 2500 at R = 0.3M, 1300 at R = 0.6M, 
     |  :     : a|                          800 at R = 1.0M
     |  :     : n|
     | red    : .|
     -------------

FIGARO IMAGE displays flips CCD images from top to bottom.



Helen Davies