|
Contents
- The Device
- Use with
AAO Instruments
- UCLES
- UHRF
- Others (including decommissioned
instruments like Taurus and RGO)
This page contains information on
the AAO's science grade EEV 2Kx4K device
- hereafter EEV2. It was first commissioned on the AAT in April
2001.
The information provided is based on tests made in at the Epping Lab by
John Barton (a copy of his comprehensive report can be found here),
and observations made at the AAT with the RGO Spectrograph by
R.Stathakis, and UCLES by S.Ryder.
1. The Device
- Identity & Format
The CCD array has a 2048
(Horizontal) x 4096 (Vertical) pixel format (with
an additional 4 unused ertical pixels). Each pixel is 13.5 x 13.5
micrometres
in size, for a total active image area of 27.6 x 55.2 mm. The device
has
been thinned, and is used back-side illuminated. Its performance has
been
optimised for blue wavelengths.
This is the grade 1 (science grade) contract device (graded to Paul
Jorden's revised specfications), received from RGO on 10/11/98. It has
serial number 7461-16-7 and A6284 marked on the CCD header.
- Binning
- Binning in possible in the the vertical or horizontal
directions.
- Quantum Efficiency
- We currently have no direct measurements of QE ourselves
with this device.
The QE as measured by EEV is shown below. These numbers would make it
by
far the AAO's most blue sensitive detector.
| Wavelength (nm) |
300 |
350 |
400 |
500 |
650 |
900 |
| Qauntum Efficiency (%) |
25 |
35.9 |
73.2 |
91.7 |
86.2 |
30.1 |
 |
A larger version
of this figure is also available, as is a colour
postscript file. The data used to produce it can be found here. |
NB: because the Postscript file was created by an
Evil
Empire Microsoft product, it may not preview with
Ghostview version 1.5 or earlier. It will preview with Ghostscript
version 4 or later, and it will print out.
|
- Limited comparisons of RGO spectrograph performance with
the
EEV and archive TEK data have been obtained. We compared data for
the standard LTT4364 through wide slits taken with the RGO centred at
6581A,
1.7A/pix, with archive TEK data with the RGO centred at 4778A,
3.0A/pix.
Unfortunately, althrough the EEV data extends over the full wavelength
range covered by the TEK data, the ends of the EEV data are
significantly
vignetted by the spectrograph camera. So we only get a decent
comparison
over the wavelength ranges of the EEV corresponding to similar
physical
locations in the spectrograph focal plane as the TEK - roughly
4000-6000A.
The figure below summarises the
photons/unit wavelength/time comparison.
From 4000-6000A the EEV is 30-40% better than the TEK, as expected from
the manufacturers numbers above. So although we don't have confirmation
of the EEV's excellent blue response, we have no reason to disbelieve
the
manufacturers numbers. They are at least confirmed from 4000-6000A.

- Fringing
Fringing with this device is much
worse than seen in the MITLL large format
CCDs, and even worse than that seen in the TEK CCD.
Observations with the RGO spectrograph indicate spectroscopic fringing
at a level of <0.5% p-p at 6000A, 0.5% p-p at 6500A, 4% p-p at
7000A,
22% at 8000A , 58% p-p at 9000A and 60% at 10000A.
This compares with 3% p-p at 8500A for the MITLL3, and 6.5% p-p for
the MITLL2a.
The following GIF images shows a
10-pixel cut averaged down the centre
of the EEV2 CCD, and very roughly normalised as a function of
wavelength
using a spline. The big wiggles near 6000A are not fringing - they're
just
my spline not matching the flat field's blaze function very well.
- Cosmetics
The device has a number of hot
pixels (see John
Barton's report). The hottest pixel on the device at (1842,1525)
generates
a faint hot column of about 10 e-/pix as the rows are shifted through
the
hot pixel during the readout. Thus the hot column appears only on the
"far"
side of the hot pixel and extends right through to the last row of the
readout. No other hot columns have been detected.
Flat field cosmetics appear to be very good. Anecdotal reports from
the first spectroscopic commissioning run suggest a virtually
featureless
flat-field in the blue. Its hard to tell what the flat field
performance
is like in the red, as the observed flat fields are completely
dominated
by fringing effects, giving a dominant 'diseased zebra' pattern.
The 'diseased zebra' : the image below shows a sample
914x150 pixel
spectroscopic flat-field sample. The left edge is ~8500A, and the right
edge ~10100A at 1.7A/pixel. The stretch is such the the features have
60%
peak-peak amplitude.
Lab results show regularly spaced lines across flat-fields.
These occur
every 512 pixels at the boundaries of the stitching process employed in
the manufacture of these large CCDs. Each line appears as a pair of
rows,
one brighter and one fainter than the average for the flat field. The
worst
pair exhibits a row about 1% brighter and the next row about 4% fainter
than the average, i.e. a 5% p-p effect. Hopefully these will flat-field
out. This worst pair, in fact, form a limit to the full well for the
CCD
as this pair had a weaker charge handling capacity than the remainder
of
the CCD.
There are no light emitting defects. There are several
trapping sites on this CCD but they are relatively
small. See John Barton's report
for
more details.
- Cosmic Rays
- A quick visual examination of an 1800s dark count image
shows about 400
CR hits per 1800s exposure over the whole detector (where a 'hit' is a
several hunred adu peak CR). This is much lower than the MITLL3
(about 1500 hits in the same time, though the detector has a 23% larger
surface area).
- Focus
- The EEV detector sits in a standard AAO dewar, and will
mount on all the
AAO instruments - however it has a preferred orientation because it is
rectangular. This means, for example, that on the RGO it has to be
mounted
???? relative to the TEK.
- Performance
| SPEED |
INTEGRATION
(us) |
GAIN
(e-/adu) |
READNOISE
(e-) |
ALPHA
(x1e-6) |
SAT
(Ke-) |
READ
RATE
(us/pix) |
Read Time (s) |
| Full |
2x2 |
5x5 |
| NONASTRO |
1+1 |
6.8 |
8.2 |
1.16 |
200 |
7.5 |
69 |
30 |
11 |
| FAST |
2.5+2.5 |
2.7 |
4.4 |
0.59 |
168 |
11.5 |
105 |
43 |
54 |
| NORMAL |
5+5 |
1.3 |
3.4 |
0.35 |
81 |
16.5 |
150 |
54 |
17 |
| SLOW |
20+20 |
0.32 |
2.7 |
0.11 |
21 |
46.5 |
416 |
123 |
31 |
Readout times include a reasonable overscan. Full
is 2120 pixel
(incl 72 pixel overscan) by 4096 rows unbinned. 2*2bin is 1060
binned
(by 2) pixels (incl 48 overscan bins) by 2048 binned (by 2) rows. 5*5bin
is 424 binned (by 5) pixels (incl 14 overscan bins) by 420 binned (by
5)
rows.
- Performance on the Telescope
- The performance figures above were derived in the lab at
Epping by John
Barton. Som eo fhtese have been checked on the AAT in a commissioning
run
April 1-2, 2001 with the RGO spectrograph. Using the IRAF findgain task
in a window [951:1150,2000:2100] of the full EEV CCD we confirm the
following
read noise and gain numbers, which give us confidence that the
performance
table above has been verified on telescope. The pairs of numbers are
completely
independent measurements from seperate bias and flat field frames.
| SPEED |
GAIN
(e-/adu) |
READNOISE
(e-) |
| FAST |
2.60,2.57 |
4.64,4.57 |
| NORMAL |
1.29,1.29 |
3.48,3.36 |
| SLOW |
0.29,0.32 |
2.36,2.62 |
On the telescope the dark current in unbinned mode has
been measured
as 1.2+-0.1 and 0.65+-0.1 e/pix/1800s in a pair of consecutive
exposures
at the start of a run (the decrease is probably due to settling after
power
on as described from lab tests below).
Linearity tests in speed FAST confirm that the linear
non-linearity
correction alpha is actually a good approximation for this CCD (at
least
in speed fast, and probably in the slower speeds). (This is not
true
for the MITLL CCDs where higher order corrections than alpha are needed
- see the WFI
linearity discussion for more details. )
The coefficient measured (+0.59e-6) is consistent with
JRB's lab result
in the table above, so these can be used with some confidence by picky
observers.
- Dark Current
- Dark current is only detectable in binned images and then
only if account
is taken of the clock induced dark currnet - the latter dominating in
short
exposures. The dark current generation rate is about 0.15
e-/pix/2000sec
or 0.27 e-/pix/hour. On the telescope the dark current in unbinned mode
has been measured as 1.2+-0.1 and 0.65+-0.1 e/pix/1800s in a pair
of consecutive exposures at the start of a run (the decrease is
probably
due to settling after power on as described from lab tests below). This
is negligible for most applications (unless exposures are longer than
500
seconds and a binning of 8 or more pixels is used).
Although only detectable in binned images (eg 5 x 5 binning), clock
induced dark currents are comparatively strong on this CCD. The charge
build-up is uniform across the CCD and independant of exposure time. It
amounts to about 0.10 to 0.12 e-/pix, so is negligble unless binning
more
than 20 pixels together. It dominates shorter exposures if the CCD is
properly
cleaned out, i.e. not recently powered on.
This CCD recovers rapidly after power-on compared to
others. After 1
hour the dark current rate is down to about 1 e-/pix/2000sec compared
with
the MITLL CCDs which require 8 hours to get to this level. The CCD
cleanout
rate following power off and on, with a cold CCD, as measured in dark
frames
are:
|
Time after Power on
|
Intensity
(e/pix/2000s)
|
|
30s
|
123
|
|
5min
|
12
|
|
10
|
6.3
|
|
20
|
2.9
|
|
30
|
1.3
|
|
60
|
1.2
|
|
80
|
0.7
|
|
100
|
0.5
|
|
16 hour
|
0.35
|
- RESIDUAL IMAGES
- For the EEV, residuals from non-saturated images or from
saturating overexposure
should not be a problem.Tests were made for the residual cleanout rate
by overexposing a small area of the CCD by a factor of 10 times
saturation
level, i.e. 2Me-/pix, and then taking a series of 100 sec exposures
starting
30 secs after the end of the saturating exposure. This test indicated
that
the residuals accumulated at a rate less than 1 e-/pix/2500sec after 30
seconds and at 4 minutes were undetectable in a 5*5 binned image.
3. Use of the
Device on AAO Instruments
3.2 UCLES
The EEV2 device was commissioned
with UCLES in early July 2001 by Stuart
Ryder (sdr@aaoepp.aao.gov.au).
It should be the detector of choice for most programs when working
blueward
of H-alpha (6563 A). At longer wavelengths, the significant fringing of
the EEV could be a problem, and the MITLL3 has better
QE in any case.
Windows
The same readout windows
used for the MITLLL2A/3 detectors
can also be used with the EEV2, as the array dimensions are the same.
The
smaller pixels give improved spatial and spectral resolution cf. the
MITLL
devices, at the expense of 10% less coverage of the chellogram. The
EEV2
device shows no sign of the charge diffusion, which degrades the
maximum
resolution of UCLES with the MITLL2a (R<70000 for even 1 pixel
slits).
By closing the slit down to 0.3", a resolving power R=115000 at 5500 A
is attainable with the EEV. More realistically, a 0.5" slit will yield
R=85000 (though slit losses due to seeing will still be significant).
Format
When viewed on the XMEM
display, EEV2 images have red (lower
number) orders on the left, and wavelength increases going up each
order.
When saved to disk and viewed with figdisp/Ximtool, the images are
flipped
vertically, so that wavelength increases from top to bottom along each
order. This is the same orientation as the Tek, and the exact reverse
(ie
flipped in X and Y) of the MITLL2A/3 orientation.
Dark current
Just 36 hours after being
installed, the dark current stabilises
to a level of 0.9 e-/pix/hr. This is rather better than the MITLL2A can
reach in the same time, but still higher than the lowest values reached
in the lab. The difference is due to a faint glow from the coude room
walls,
which we hope soon to reduce by re-painting the walls.
Fringing
Fringing is apparent with
the EEV2 at wavelengths beyond 6000 A, and becomes worse as one
goes
further into the red. At 6000 A, the fringing is at a level of <2%
p-p;
it is still <3% p-p at 6500 A, but rises to 6% p-p at 7000 A, and
36%
at 8000 A.
However, careful flat-fielding can calibrate this;
for instance, dividing the spectrum of a standard star by a flatfield
image
taken immediately afterward (see figure below) reduces the fringe
amplitude
at 7500 A from 20% p-p, to 4% or less. Thus, unless the wavelengths of
primary interest are redder than 6500 A (in which case the MITLL3
should
be favoured due to its superior quantum efficiency), the fringing
should
not be a serious hindrance to most UCLES programs.
Observing
Observers are reminded to
enter the command "DATA USHORT" after
running up the CCD system in order to keep the file sizes manageable.
An important point to note is that the current UCLES
camera optics
cannot illuminate the entire area of the EEV2 detector (27.6 x 55.2
mm).
In fact the unvignetted (<10% vignetting) region which can be
observed
is more like 18.8 x 38.5 mm. The region covered at 50% vignetting is 60
x 34 mm, which is approximately the entire chip, however unless you are
working in the very red, the echellogram will not put any light on much
of the chip.
You can see the effect of this by loking at the figures on
the MITLL3 and MITLL2a
pages, but note that the EEV
CCD is physically
10% smaller than these devices, so it sees 10% less of the echellogram
than
the figures indicate, and is therefore less subject to vignetting.
3.3 UHRF
The EEV2 device is now the standard
large format device used on UHRF (MITLL3 is not generally useful, since
it cannot be binned).
3.4 Others
Taurus
MITLL3 has been used with
Taurus in the past. Taurus is now decommissioned for use on the AAT.
This information is provided for historical completeness only.
Contact Joss Hawthorn ( jbh@aaoepp.aao.gov.au)
| Field Sampling with Taurus II & EEV2 |
|
Detector |
f/8 |
f/15 |
| TEK |
0.594"/pix |
0.315"/pix |
| MITLL2a & MITLL3 |
0.37"/pix |
0.20"/pix |
| EEV2 |
0.33"/pix |
0.18"/pix |
The MITLL detectors allow an 8.4'
(vertical) x 9.5' (horizontal) field
to
be charge shuffled and stored side by side on the detector. This has
important uses for Taurus++ slit observations, and TTF imaging with
band switching.
The EEV2 detector allows an 7.5'
(vertical) x 9.5' (horizontal) field
to
be charge shuffled and stored side by side on the detector. This has
important uses for Taurus++ slit observations, and TTF imaging with
band switching.
RGO Spectrograph
MITLL3 has been used with RGO in the
past. Taurus is now decommissioned for use on the AAT.
This information is provided for historical completeness only.
Contact Chris
Tinney
The EEV2 device has been used on
RGO spectrograph's 25cm camera. RGO
users get smaller pixels than the MITLL devices (13.5um vs 15um), and
superior
blue sensitivity. Once again vignetting stops the whole chip from being
used, with only about 3500 pixels being illuminated by the RGO. (This
is
still more than double the wavelength range covered by the TEK1K). You
can play with the possible options using the WWW version of RGOANG.
|