The condition for maximal transmission of light with wavelength
through a Fabry-Perot interferometer is
The ratio
characterises the relative influence of phase
change at a given wavelength. Combining Eqns. (9) and
(10) for a common wavelength and plate spacing L, gives
Figure 5(a) shows four TTF scans at the lowest
plate spacings reached by our instrument. The scans show
blended lines of Ne (659.9, 667.8 and 671.7 nm). The lines
are unresolved at all plate spacings except the largest (L = 4.7
m),
where they are labelled in (a).
The scans were made at various values of
(Zc, Zs) and
transformed to physical units of spacing by Eqn. (2).
Changes in the software scan increment (
)
are evident
in the different sampling densities of each scan. Observe that the transmission
peaks are evenly separated by
m,
confirming that the calibration in Eqn. (2) is robust over
all settings of Zc and Zs used. Also note the broadening of the
transmitted profile as plate spacing and resolution decreases.
The flat background levels are from CCD regions that were not used
in the charge shuffle.
In Fig. 5(b) we plot the change in
for the same orders of the Ne blend shown in (a).
The ratio
was calculated at each plate spacing
measured in (a) by Eqn. (11).
Queensgate Instruments have measured
for TTF at 666 nm.
Observe in Fig. 5(b) that phase correction
is a
% effect at 670 nm. The dotted lines show
the effect to be significantly less at wavelengths near the centre
of the TTF coverage.
The narrowest spacing (2.5
m) is a self-imposed limit to which we
are prepared to
drive the plates. Any closer and we run the risk of damaging the inner
coatings by pressing dust particles between the two coating surfaces.
We conclude that at the narrowest spacings of TTF
we are in a regime where phase effects are non-negligible,
particularly for wavelengths at the extremities of the optical coating curve.