By all means read this, but note that
it has been replaced by a set of systematically organised
manual pages.
This is intended as an aide de memoire in the development of
the PN Spectrograph and as a check that all operational aspects
hang together coherently. It is not intended as an imperative
document but to stimulate discussion.
The PN Spectrograph
The instrument will consist of:
Storage
The PN Spectrograph will be stored as a complete unit with the optics in
place (but with [OIII] detectors removed) either at the WHT or at the TNG
(Checklist 6). The Halpha camera will remain attached (assuming that it exists).
It will need to be moved from one dome to the other up
a steep hill and attached to the appropriate telescope.
Interface
Data will be collected via the data acquisition of each telescope
since the [OIII] cameras will belong to the respective observatory.
It may be that the Halpha camera has its own controller and data
storage facility.
The interface box will house power drivers for the
actuation of the shutter (unless
we use small iris shutters attached to the dewars, in which case
they can be run from the observatory), and the calibration matrix.
Calibration lamps, on the other hand, probably belong to the telescope.
Assuming that the two systems do NOT have an interface,
a typical calibration observation might consist of:
With an analogous procedure for observing the target.
Focus
Filters
Can only be changed off-line, and cannot be removed from the beam
during the night. To cover a range of 67Å, we will have three
filters: 5001/36, 5023/36, 5046/36. From experience the flat top
on these filters will have a width of about 0.63 FWHM or 22.3Å.
Calibration
For our purposes we would like to have three lines
within any given filter bandpass in order
to get a reliable and unambiguous dispersion solution.
At the WHT, CuNe and CuAr lamps can be used. Here is the
relevant part of the combined spectrum (the major lines in the
5000-5050 part of the spectrum from L to R are:
He and Ar lamps together have the following emission lines
(of which some are very faint):
The matrix mask will include both pinholes and short slits
perpendicular to the dispersion direction. When illuminated
by an arc lamp, these will provide an excellent check on image distortion
and on dispersion over the field. It should be done while the telescope
is pointing near the target, so avoid effects caused by differential
bending when the telescope is slewed.
This process will provide a well-defined zero point for the
velocity/offset relation between the two arms, when two arms are in use.
If it is performed carefully it will also give the zero point in one-arm
operation.
To correct for small shifts in the image centre between integrations
(recall that there may be no easily-visible point sources with which
to align individual frames) we should be able to use the dispersed
images of foreground stars, which look like trails
(Ken has tested this). In any case, it
may be useful to reobserve the matrix mask
at regular intervals.
We have
no totally reliable method of correcting for telescope errors except for
the star-trail method.
KEN FREEMAN'S NOTES (email 29/7/98):
ALIGNING IMAGES WITH THE FILTER PROFILE. This worked really well, even with
my filter which does not have very sharp cutoffs. I guess it works because
the filter is typically 40A wide and we need to align to about 0.2A. There
is plenty of signal in the stellar spectra, so this is not so demanding.
It needs corrections for the color of the individual stars (derived from
the spectra themselves) - this correction is easy and small. I also had to
correct for variations of the passband over the filter - this was measured
from the matrix mask data and is a problem that will be *much* smaller with
the PNS - my filter is not so good and was near the focal plane, while for
the PNS the filter is near the pupil. Despite all this, I recovered the
absolute velocities of galactic PN to 15 km/s.
ILLUMINATING THE FIELD WITH WHITE LIGHT. This is partly for flatfielding
without the matrix mask (we have the filter in there to reduce the bandwidth),
and partly with the matrix mask for registration as discussed above. The white
light observations through the matrix mask are used to measure variations of
the passband over the field. We could manage with dome flats though something
a bit hotter than the usual dome flat lights would be nicer. But we do need
a good source of He illumination for the matrix mask. John thinks this doesn't
contribute much to the cost, and it would be only a small thing to add our
own hot white light source as well.
return
sequence PNS CONSOLE TELESCOPE CONSOLE
1
2 Switch on lamps
3 Engage matrix mask
4 Clear CCD (readout)
5 Open shutter 10s
6 Readout CCD
5009.3
5017.2
5031.4
5037.8
4921.929 HeI
4965.0795 ArII
4972.1597 ArII
5009.3344 ArII
5015.6779 HeI
5047.738 HeI
5062.0371 ArII
The following lines are medium-strength
in the (UES) Th-Ar catalogue:
4985.372
5002.097
5009.334
5017.163
5028.636?
5039.23
5044.719
5047.043
5050.784
5062.037
Possibles from the Al/Ca/Mg-Neon catalogue:
4957.033
4973.538
5005.16
5031.3484
5037.7505
5041.620